Straightness
Reconstruct a guideway profile from small angular changes.
Laser source instead of halogen: stable return beam over long distances and wide ranges.
A collimated diode laser illuminates the mirror; a high-resolution detector reads the return spot. Some designs use polarising beamsplitters and quadrant detection to stabilise the zero.
Standard numbers for checking method clauses. Full texts are not redistributed; listing does not certify an instrument.
Reconstruct a guideway profile from small angular changes.
Define the datum before evaluating a nominal right-angle relationship.
Compare small angular movements and precision indexing positions.
Check direction and lateral position separately during optical assembly.
Level the machine into a measurable state before establishing an angular datum.
Use the line through the first and last points; sum the extreme deviations.
Adopt the line or plane minimising the sum of squared residuals.
Measure the same line by two independent principles and compare profiles.
Build a line of sight with a telescope and targets as a datum axis.
Use angle gauges as material standards and read the difference with an autocollimator.
Derive the wedge from the angular separation of the two face reflections: α=Δθ/(2n).
Measure a long guideway in sections and splice them via overlaps.
Measure column guide-face deviations from vertical against a level datum.
Verify indication errors across the range with an angle generator or table and polygon.
Our catalogue focuses on angular metrology; this technique is listed for method comparison and system planning.