Principle
The calibration feeds the autocollimator known angles and compares indications with the standards. Two standard families are common: small-angle generators (flexure-based sine mechanisms or interferometric angle rigs) producing precisely known angles near zero, or a polygon on a high-accuracy rotary table producing standard rotations around the circle. A mirror on the moving part of the standard carries the angles while the autocollimator reads them; the differences are the indication errors. The range is sampled at equal steps (e.g. every 100″ or 1′ depending on range and class) in both directions to assess hysteresis and drift.
Preparation
Confirm the standard's traceability and uncertainty, which should be better than the unit under test, typically by a factor of three or more (rule of thumb). Mount the mirror on the moving part of the standard, normal to the autocollimator. Warm up the instrument for the specified time and record temperature. Plan the calibration points to cover the range and the working interval.
Workflow diagram
Procedure
- Set up the standard with confirmed traceability; mount the mirror on the moving part and align it to the autocollimator.
- Warm up the instrument, record the zero reading and begin.
- Present the standard angles point by point as planned, recording the autocollimator indication.
- Repeat the points in reverse after completing the forward run to check hysteresis and zero drift.
- Compute indication errors e=indication−standard angle and, if needed, fit an error curve for interpolated corrections.
- Report per-point errors or corrections, forward-reverse agreement, zero drift and calibration uncertainty.
Calculation
Per-point error e_i=R_i−S_i, with R_i the indication and S_i the standard angle. The forward-reverse difference h_i=|e_i(fwd)−e_i(rev)| indicates hysteresis. The correction is c_i=−e_i, added to indications in use. Zero drift d=final−initial zero reading; data from out-of-tolerance intervals are rejected or re-measured. A combined uncertainty is u(e)≈√(u_S²+u_R²), u_S from the standard and u_R the repeatability (simplified example).
Illustrative example
Teaching data, not measured: at the +500″ point an autocollimator shows errors of +0.4″ forward and +0.2″ reverse (demonstration), with a standard uncertainty of 0.1″ (k=1, example). The correction at that point is −0.3″ using the mean of both runs; the report also includes the 0.2″ hysteresis and zero drift. No real instrument is implied.
Limitations
- Calibrating only near zero leaves the working interval's errors uncovered.
- A standard with uncertainty comparable to the unit under test voids the calibration.
- Unrecorded hysteresis and zero drift make corrections ill-suited for reverse travel.
Reporting
Retain raw data, datum definition, temperature, repeatability, algorithm and uncertainty. Illustrations are not test results.