MICROCRE / OPTICAL METROLOGY400-812-7833  ·  SHANGHAI
研润 WHW
MICROCREPRECISION IN EVERY ANGLE
Home / Establishing a Datum Line with a Collimator Telescope
MEASUREMENT METHOD

Establishing a Datum Line with a Collimator Telescope

Establish a straight datum spanning the full length of a machine tool, assembly platform or large structure to measure position deviations of bores, guideways or components from that axis. It suits coaxiality and alignment over long distances with many stations. It is often paired with an autocollimator: the telescope provides the position datum, the autocollimator the angular datum.

Principle

A collimator telescope sights a concentric target such as a glass target with a centred pattern; the offset between the reticle cross-line and the target centre converts to lateral displacement through the objective imaging relation. The datum line joins the targets at two reference points, analogous to an endpoint reference, and offsets of intermediate targets give their position deviations. In autocollimation mode a mirror in front of the objective also yields its tilt about the optical axis, so position and angle are measured simultaneously.

Preparation

Check the telescope's focus-induced line-of-sight error over the working distance and correct per the manual if needed. Match targets and mounts to the bores or datums and verify the centring of target versus mount rotation axis. Set the two reference points rigidly so they do not move mid-measurement, and record the thermal environment.

Workflow diagram

01Set telescope and datum targets02Focus and verify sight line03Measure intermediate offsets04Convert to deviations05Report datum deviations06Assemble and report

Procedure

  1. Mount targets at the two reference points, install the telescope and roughly align the sight line through both datum targets.
  2. Focus stepwise to each working distance and record focus-induced sight-line shifts.
  3. Define the datum by zeroing on the two datum targets (endpoint convention) and read the offsets of intermediate targets from the cross-line.
  4. Convert offsets to linear deviations with the imaging scale, separately in the horizontal and vertical directions.
  5. Where angular data are needed, read the mirror tilt to the sight line in autocollimation mode.
  6. Assemble position deviations and angles, and report the datum definition, environment and uncertainty.

Calculation

Target offset conversion: Δ=K·d, where d is the micrometer reading difference and K the conversion factor at that distance from the manual or calibration. Deviations follow the endpoint rule: ei=Δi−[Δ1+(Δn−Δ1)·(xi−x1)/(xn−x1)]. Angular readings convert as in autocollimation: θrad=θ″·π/648000.

Illustrative example

Schematic · not measured data
Schematic: telescope sight line and target positions; readings illustrative.

Teaching data, not measured: two datum bores 2 m apart with a measured bore at 1 m. With both datum targets reading zero, the intermediate target converts to +0.03 mm horizontally, i.e. a +0.03 mm horizontal deviation from the datum line. The numbers only illustrate the conversion, not any instrument performance.

Limitations

  • Unverified focus error bends the sight line and biases readings at different distances.
  • Eccentricity between target centre and mount rotation axis introduces an offset.
  • Datum points displaced by force or temperature mid-measurement invalidate the whole line.

Reporting

Retain raw data, datum definition, temperature, repeatability, algorithm and uncertainty. Illustrations are not test results.

Method reference ↗ Method reference ↗ 
STEP BY STEP

Understand the measurement step by step

Principle schematic · not to scale
1 / 5
  1. Wide view of the beam with targets at both ends and the telescope set up, the line of sight through both centres.
  2. The cross-hair steps through the five stations while the offset appears in the corner of the screen.
  3. A reflector on target B drives the autocollimator from 0.0″ to +0.4″.
  4. The datum line rises slightly and the per-station corrections are distributed by distance.
  5. The corrected offsets land, the end-point line is drawn and the 282 μm band appears.

Teaching animation: follow the stations and reference directions. Do not read dimensions or angles from the schematic.

WORKED EXAMPLE

Straightness of five bearing-seat faces on an 8000 mm gantry beam (telescope datum)

Teaching example with assumed data, not a customer project, instrument accuracy claim or acceptance result.

Example task

A gantry mill with an 8000 mm span carries five bearing-seat faces on its cross-beam. Before assembly a telescope establishes a stable datum line and the lateral offset of every seat is measured against it. The telescope stands beyond one end, two glass targets define the line, and an autocollimator on target B monitors how the line drifts during the run.

Setup and preparation

  • Fix a glass target with a central aperture at each end of the beam, with the target centre at roughly the height of the seat faces.

  • Set the telescope beyond end A, roughly align the line of sight through target A, then focus on target B at 8 m and centre the cross-hair on its aperture.

  • Define a station every 2000 mm along the beam and place a cross-hair target holder on each seat, pressed against the machined locating face.

  • Refocus at every station and re-check the zero of targets A and B after each focus change, recording the shift caused by focusing.

  • Tape a reflector to datum target B and read it with the autocollimator (0.01″) at the start and the end of the run to monitor datum drift.

  • Record 19.6 °C and keep the beam out of direct sunlight.

Example data and records

  • Datum target A (0 m) zeroed: 0.00 mm

  • Datum target A (0 m) zeroed; station 1 (2 m) telescope offset +0.12 mm

  • Datum target A (0 m) zeroed; station 2 (4 m) telescope offset -0.15 mm

  • Datum target A (0 m) zeroed; station 3 (6 m) telescope offset -0.08 mm

  • Datum target A (0 m) zeroed; station 4 (8 m) telescope offset +0.05 mm

  • Datum target B (8 m) re-sighted after the run; the autocollimator shows +0.40″ drift, equivalent to 15.51 μm at the far end

  • Telescope micrometer division 0.02 mm, estimated to 0.01 mm; cross-hair re-checked after each focus change

  • 19.6 °C, no direct sunlight, beam freely supported

Calculation steps

The datum line joins the centres of targets A and B. At the end of the run the autocollimator shows target B +0.40″ away from its start position, which is 8000×tan(+0.40″)=15.51 μm at the far end, so the datum tilted slightly during the run. Distributing that tilt linearly gives corrections cᵢ=+0.0000 → +0.0039 → +0.0078 → +0.0116 → +0.0155 mm and corrected offsets +0.000 → +0.116 → -0.158 → -0.092 → +0.034 mm. Evaluated against the end-point line, with +8.6 μm of correction per station, the residuals are +0 → +107 → -175 → -118 → +0 μm. The largest positive deviation 107 μm is at station 2 (2 m) and the largest negative -175 μm at station 3 (4 m), giving a straightness of 282 μm.

Interpretation

Straightness of the five seats against the datum line: 282 μm (order of 0.28 mm). This passes the 0.3 mm tolerance for a beam seat face with 18 μm margin; the drift correction of only 15.5 μm does not change the verdict.

Verification and pitfalls

  • Datum drift must be monitored and removed: 15.51 μm here, and ignoring it would shift the middle stations by about 7.76 μm.

  • Re-check the zero of both targets after every focus change; telescope focusing shifts the line of sight and that error goes straight into the readings.

  • The target holder must be pressed against the machined face; a magnetic stand or a hand-held holder introduces a tilt that grows into millimetre-level error.

  • Nobody may lean on the beam or the tripod during the run: at 8 m a 1″ line rotation is 39 μm.

  • One-sided sunlight bends the beam, so schedule the run away from direct sun and record the temperature.