Home / Measuring the Wedge Angle of an Optical Parallel
MEASUREMENT METHOD

Measuring the Wedge Angle of an Optical Parallel

Measure the small wedge between the two surfaces of optical parallels, glass plates or plane-parallel windows. The wedge affects the parallel's use as a length and flatness standard and the beam walk-off in transmission. It suits fast non-contact measurement of individual parallels in a metrology room and incoming inspection.

Principle

When an autocollimator sights a parallel, the front and rear surfaces each form a reflection; the angular separation of the two reticle images depends on the included angle and the index. Refraction and reflection give the rear-surface image an offset of 2n·α from the front-surface image (n the refractive index, α the true wedge), hence α=Δθ/(2n) with Δθ the image separation. The image-pair orientation gives the wedge direction. Anti-reflection coatings dim the rear image but it remains measurable; slight rear masking can suppress stray reflections.

Preparation

Clean both faces, ensuring freedom from fingerprints and particles. Rest the parallel on a stable support normal to the autocollimator line. Confirm both reflection images are visible and their separation lies within range (an over-large wedge leaves one image or exceeds the range). Record the refractive index and its temperature coefficient.

Workflow diagram

01Clean and mount02Find both images03Read separation04Look up index05Compute wedge and direction06Re-measure and report

Procedure

  1. Mount the cleaned parallel and focus the autocollimator on a sharp surface reflection.
  2. Adjust the incidence so the front-surface image lies near the field centre.
  3. Locate the dimmer rear-surface image and confirm it is fully readable.
  4. Read the separations Δθx and Δθy between the images in two orthogonal directions with the eyepiece micrometer.
  5. Compute the wedge α=√(Δθx²+Δθy²)/(2n); the image-pair direction gives the wedge orientation.
  6. Re-measure at a different orientation to confirm stability and report wedge, direction and uncertainty.

Calculation

Separation Δθ=√(Δθx²+Δθy²) in arcsec; true wedge α=Δθ/(2n) with n the index at the measurement temperature (about 1.5163 for BK7-type glass, example). The wedge direction is β=arctan(Δθy/Δθx). Example: Δθ=30″ and n=1.5163 give α≈9.9″ (teaching demonstration).

Illustrative example

Schematic · not measured data
Schematic: separation of front and rear reflections and the wedge direction; values illustrative.

Teaching data, not measured: image separations Δθx=28.8″ and Δθy=8.4″ give Δθ=30″ (demonstration). With n=1.5163, the wedge is α≈9.9″ at β=arctan(8.4/28.8)≈16.3°. The numbers only demonstrate the conversion.

Limitations

  • Mistaking 2n·α for α or n·α halves or mis-scales the result.
  • Misidentifying the rear image as a stray front-surface image reads the wrong separation.
  • An index mismatched to material or temperature biases the conversion.

Reporting

Retain raw data, datum definition, temperature, repeatability, algorithm and uncertainty. Illustrations are not test results.

Method reference ↗ Method reference ↗ 
STEP BY STEP

Understand the measurement step by step

Principle schematic · not to scale
1 / 5
  1. The plate stands on its support and two cross-hair images, one bright and one dim, appear in the field.
  2. The front image is moved to the field centre and the cross-hair is aligned.
  3. The back image sits 7.2″ off in x and 4.3″ in y, and the micrometer reads both.
  4. A line joining the images is drawn with the azimuth and separation alongside.
  5. α=Δθ/(2n) gives 2.76″ and converts to 0.67 μm of thickness difference.

Teaching animation: follow the stations and reference directions. Do not read dimensions or angles from the schematic.

WORKED EXAMPLE

Parallelism (wedge angle) of a plane-parallel plate

Teaching example with assumed data, not a customer project, instrument accuracy claim or acceptance result.

Example task

A shop check on a 20 mm thick plane-parallel plate with a 50 mm clear aperture. The front and back surfaces each return an image, and the micrometer eyepiece of an autocollimator (0.01″) reads their separation in x and y. The true wedge angle follows from α=Δθ/(2n), and the thickness variation over the 50 mm aperture is derived from it.

Setup and preparation

  • Stand the plate on three-point supports, clean both faces and confirm there are no fingerprints or dust; no tape on the back surface.

  • Aim the autocollimator at the plate centre and focus on the front-surface image first, then fine-focus to find the weaker back-surface image.

  • Trim pitch and yaw until the front image sits near the field centre and both images are fully readable.

  • Read the x and y separation between the two image centres with the micrometer eyepiece, three times each, and average.

  • Rotate the plate 90° about the beam axis and repeat, to separate true wedge from surface-slope contributions.

  • Record 20.0 °C and avoid touching the plate edges during the run.

Example data and records

  • Front-surface image centred in the field: (x, y)=(0.0″, 0.0″)

  • Back-surface image displaced from the front image by +7.2″ along x

  • Back-surface image displaced by +4.3″ along y

  • Plate rotated 90° about the beam axis: x=4.2″, y=−7.3″, separation still 8.4″

  • Plate thickness 20.00 mm, clear aperture 50 mm, index taken as n=1.5168 for BK7

  • 20.0 °C; the back image is about 60 % as bright as the front, no AR coating

Calculation steps

The image separation is Δθ=√(Δθx²+Δθy²)=√(7.2²+4.3²)=8.386″, at an azimuth of 30.8° from x. The back image is offset from the front by 2n·α, so the true wedge angle is α=Δθ/(2n)=8.386/(2×1.5168)=2.764″. Over the 50 mm aperture this is Δ=50×tan(2.764″)×1000=0.67 μm of thickness change between the two edges. After a 90° rotation the separation is still 8.4″, confirming that the reading is a true wedge rather than a local surface slope.

Interpretation

Wedge angle α=2.76″, equivalent to 0.67 μm over the 50 mm aperture (about 1.1λ at 633 nm). This passes the 10″ wedge tolerance with 7.2″ margin.

Verification and pitfalls

  • Divide by 2n, not by 2: the index 1.5168 magnifies the back-image offset by 3.03, so dividing by 2 alone overstates the wedge by 52 %.

  • A 90° re-measurement must keep the separation (8.4″ here); a clear change means surface slope is contaminating the reading.

  • The back image is dim, so lower the room lighting or use a hood; otherwise the micrometer cross-hair is unreadable and the reading shifts by more than 0.5″.

  • Both faces must be clean; a single dust particle deforms the local image and fakes a wedge above 1″.

  • The wedge direction comes from the line joining the images; assembly usually needs the direction, not just the magnitude.