MICROCRE / OPTICAL METROLOGY400-812-7833  ·  SHANGHAI
研润 WHW
MICROCREPRECISION IN EVERY ANGLE
Home / Engineering workbench
CALCULATE / UNDERSTAND / RECORD

Engineering workbench

Convert, learn and prepare. Calculations run in your browser.

Arcseconds → height difference

Δh = L × tan(θ × π / 648000) × 1000

Height difference
0.484814 μm

Enter the indicated mirror angle, not the doubled ray deflection. L is the effective base span, not the optical distance to the reflector.

Profile reconstruction · learning tool

Each angle represents one span interval. Heights are integrated and a least-squares line removed. The residual peak-to-valley is not minimum-zone straightness or an acceptance decision.

Measurement route guide

Rule-based navigation through the curated material; not a live language model or an engineering approval.

REFERENCE CALCULATIONS

More reproducible calculations

Change the inputs to check a calculation. Units, formulas and assumptions are shown; results support understanding and planning.

Angular unit conversion

deg = θ / 3600; rad = θ × π / 648000

1″ = 0.000277778° = 4.848137 μrad.

Input is in arcseconds, not arcminutes.

Mirror tilt and ray deflection

β = 2θ

Mirror tilt 10″ → reflected-ray deflection 20″.

Reflected-ray deflection is twice mirror tilt. Do not halve an already calibrated mirror-angle reading.

Focal displacement to mirror tilt

θ = ½ arctan[x / (1000 f)] × 648000 / π

x=10 μm, f=300 mm → θ≈3.437747″.

Ideal autocollimation geometry x=f·tan(2θ), ignoring distortion. Use the calibrated response for an instrument.

Support span and height difference

Δh = 1000 L tan(θπ/648000)

100 mm × tan(1″) = 0.484814 μm.

L is support span, not optical measuring distance. The result is a local height difference, not whole-guideway straightness.

Linear offset and squareness angle

ε = arctan[δ / (1000 L)] × 648000 / π

δ=100 μm, L=1000 mm → 20.626481″, 100 μm/m.

Assumes two ideal line directions in the agreed plane. Not full-surface perpendicularity; datum errors are not removed.

Repeated-reading statistics

x̄ = Σxᵢ/n; s = √[Σ(xᵢ−x̄)²/(n−1)]

Five example readings: mean 1.000000″; sample s=0.158114″; u(mean)=0.070711″.

At least two independent readings under the same conditions. Sample SD, range and uncertainty of the mean are distinct quantities.

Uncorrelated uncertainty components

u_c = √Σuᵢ²; U = k·u_c

u=[0.05,0.03,0.02]″ → uc=0.061644″; k=2, U=0.123288″.

Only uncorrelated standard components expressed in the same output unit. Correlations and degrees of freedom are excluded; k=2 does not automatically ensure 95% coverage.

Uniform thermal expansion

ΔL(μm) = α(μm/m/K) × L(mm) / 1000 × ΔT(K)

α=11.5 μm/m/K, L=500 mm, ΔT=2 K → ΔL=11.5 μm.

Example coefficient is not a guaranteed material property. Assumes uniform temperature, free expansion and nearly constant coefficient; excludes thermal bending, constraints and instrument drift.