PRINCIPLE / AUTOCOLLIMATION

How an autocollimator turns angle into a reading

An autocollimator illuminates a plane mirror with a parallel beam. The returning beam is focused again by the objective, and the image position shifts with the mirror angle. Every autocollimation measurement rests on this relationship.

Read more ↗
PRINCIPLE / DUAL-AXIS

Dual-axis measurement, field of view and targets

A dual-axis autocollimator reads two angular directions in one setup and reduces repeated adjustment. Field of view, resolution and working distance, however, trade off against each other.

Read more ↗
PRINCIPLE / ACCURACY

Reading resolution, indication error and repeatability

Resolution is usually the first number on a datasheet, yet usability depends on indication error within a stated range and on repeatability. The three are not interchangeable.

Read more ↗
PRINCIPLE / DISTANCE

Working distance, focal length and aperture

Maximum working distance is not a single figure. It follows from return-light energy, spot size, mirror aperture and air conditions, and it trades off against field of view and accuracy.

Read more ↗
PRINCIPLE / OPTICS-SETUP

Optical setup, accessories and common errors

The same instrument can produce results differing by micrometres depending on the setup. Reflector, base span, cosine error and Abbe error are the four factors that must be made explicit.

Read more ↗
PRINCIPLE / ANGLE-TO-ERROR

From angle readings to geometric error

The instrument gives angles; engineering needs straightness, flatness or squareness. Between them sit four steps: span, integration, datum and evaluation algorithm.

Read more ↗
PRINCIPLE / LEVEL-VS-AUTOCOLLIMATOR

Electronic level vs autocollimator: same angle, different jobs

An electronic level references gravity and reads tilt against the horizontal plane; an autocollimator references its optical axis and reads mirror angle against that axis. The datum difference decides which instrument fits a guideway job.

Read more ↗
PRINCIPLE / POLYGON-TRACEABILITY

Polygons: dividing the circle into a traceable angle standard

A polygon exploits circle closure: the n face-normal angles must sum to 360°, so the polygon calibrates itself without a higher-order angle standard. It is the key artefact for tracing rotary tables and goniometers.

Read more ↗
PRINCIPLE / FIELD-APERTURE-RANGE

Field of view, aperture and range: three specifications that constrain each other

Field of view sets how much angular change the instrument tolerates, aperture sets how much return light it collects, and focal length drives both resolution and size. Read the three together; optimising one alone usually fails.

Read more ↗
PRINCIPLE / ENVIRONMENT-ERRORS

Temperature, air turbulence and vibration: the three enemies of long-range angle measurement

0.1″ is about 0.5 μrad, a 5 μm displacement at 10 m. At this scale, room temperature gradients, air conditioning drafts and floor vibration can bury the signal. Environmental control is part of the measurement plan, not an accessory.

Read more ↗
PRINCIPLE / SIGN-CONVENTION

Sign and zero: where the plus and minus come from

The same data gives an inverted profile under a different sign convention. Write down the positive direction, zero and viewing direction before measuring.

Read more ↗
PRINCIPLE / INTEGRATION-CLOSURE

Integrating angles: step, span and closure

Angle to height is geometry, not a rule of thumb. Mismatched step and span, or excessive closure, produce a plausible but irreproducible profile.

Read more ↗
PRINCIPLE / REFRACTIVE-INDEX

Air refractive index: the hidden scale

An interferometer uses wavelength as its scale, and wavelength in air depends on temperature, pressure and humidity. Uncompensated, tens of micrometres over 10 m can come from air alone.

Read more ↗
PRINCIPLE / LEAST-SQUARES-VS-ZONE

Least squares versus minimum zone

Software defaults to least-squares residual peak-to-valley while drawings often specify minimum zone. The difference can reach 20–30%, so mixing them invalidates the decision.

Read more ↗
PRINCIPLE / LINEARITY-INDICATION

Indication error, linearity and repeatability are three different numbers

A catalogue line reading "accuracy ±1″" leaves three questions open: over which angular range, against which reference, and with what scatter. Only separate figures can be compared.

Read more ↗
PRINCIPLE / MIRROR-FLATNESS

How mirror figure error enters an angle reading

Autocollimation treats the mirror as the angle reference. Local slope adds twice to the return direction and lands directly in the reading, depending on aperture and spot position.

Read more ↗
PRINCIPLE / THERMAL-GRADIENT

Thermal gradients: why morning and afternoon disagree

Uniform heating changes size; a gradient changes shape. For angle measurement the resulting bow is worse than expansion, because no single coefficient corrects it.

Read more ↗
PRINCIPLE / SAMPLING-FILTER

Sampling, filtering and jitter: when averaging helps and when it lies

A jumping reading is not necessarily a fault. Too low a sampling rate folds fast disturbances into slow wander; too much averaging erases real steps. Both make a report look cleaner than reality.

Read more ↗