Animated explanation
- Establish the geometry
- See how one reading forms
- Change a condition and observe
- Return to the formula and its limits
The animation explains relationships and directions. Read numbers from the formula, the standard and your records.
Estimate the magnitude first
A 1 m steel part heated by 2 K grows about 23 μm.
A 1 K difference between top and bottom bows a beam in the middle while the ends stay almost fixed — the usual source of morning/afternoon disagreement.
A criterion for soaking time
Do not use "two hours" as the criterion. Use the rate: drift below one third of the target uncertainty over 15 minutes.
Log ambient, part surface and instrument temperature; disagreement among the three means a gradient persists.
What the record must contain
Start and end temperature, stability criterion, sunlight and air-conditioning exposure, and whether paired runs were made.
Related methods and industries
Recommended instruments

Dual-axis autocollimator, accuracy 0.1″, range 30 m

Dual-axis autocollimator, accuracy 1″, range 40 m