MEASUREMENT METHOD

Long-term monitoring and trending

Track accuracy over time and separate environment, outliers and real degradation.

Principle

With a stable system and fixed stations, distribution changes reflect machine state; control charts separate random from systematic drift.

Preparation

Fix stations, instrument, brackets and time window; define a baseline and change limits.

Workflow diagram

01Fix stations02Periodic re-measure03Log environment04Judge trend05Trigger maintenance

Procedure

  1. Install permanent markers and repeatable mounts at key positions.
  2. Re-measure at a fixed interval, sequence and environmental window.
  3. Log temperature, humidity, vibration and operator.
  4. Use control charts or trend lines to judge natural variation.
  5. Trigger maintenance beyond limits and keep before/after data.

Calculation

Set limits from baseline mean and standard deviation (±2σ warning, ±3σ alarm); drift rate is the trend slope.

Illustrative example

Schematic · not measured data
Schematic: baseline and control limits.

Example: baseline 4.2″ with σ = 0.6″; five points above 5.4″ indicate a systematic shift.

Limitations

  • Changing stations or brackets invalidates the trend.
  • Comparing across seasons misleads.
  • Reacting to one outlier causes over-adjustment.

Reporting

Retain raw data, datum definition, temperature, repeatability, algorithm and uncertainty. Illustrations are not test results.

STEP BY STEP

Understand the measurement step by step

Principle schematic · not to scale
1 / 5
  1. Step 1 · Install permanent markers and repeatable mounts at key positions.
  2. Step 2 · Re-measure at a fixed interval, sequence and environmental window.
  3. Step 3 · Log temperature, humidity, vibration and operator.
  4. Step 4 · Use control charts or trend lines to judge natural variation.
  5. Step 5 · Trigger maintenance beyond limits and keep before/after data.

Teaching animation: follow the stations and reference directions. Do not read dimensions or angles from the schematic.

WORKED EXAMPLE

Worked six-month drift example with 3σ control limits

Teaching example with assumed data, not a customer project, instrument accuracy claim or acceptance result.

Example task

Teaching example: one reference angle re-checked monthly, six readings, mean and standard deviation used as control limits to separate noise from trend. Synthetic data.

Setup and preparation

  • Install permanent markers and repeatable mounts at key positions.

  • Re-measure at a fixed interval, sequence and environmental window.

  • Log temperature, humidity, vibration and operator.

  • Use control charts or trend lines to judge natural variation.

Example data and records

  • Deviations in arc-seconds: +0.0, +0.2, +0.3, +0.1, -0.2, -0.4.

  • Mean −0″, sample deviation 0.261″.

  • 3σ limits: ±0.782″.

Calculation steps

s = √(Σ(xᵢ−x̄)²/(n−1)) with n = 6 gives 0.261″; limits are ±3s = ±0.782″. Seven points on one side or a monotonic run is a signal even inside the limits.

Interpretation

All points stay inside the limits, yet a one-sided climb in months 2–3 is visible; check the temperature log before calling it instrument drift.

Verification and pitfalls

  • Fix the mount, the sequence and the environmental window, otherwise trends are not comparable.

  • Keep raw readings and calibration dates; mark calibration events on the chart.

  • Derive control limits from a stable period, never from the excursion itself.