Principle
With position held fixed, time-varying readings characterize combined system and environmental stability.
Preparation
Use a fixed reflector, timestamps and temperature sensors; record power state, heat sources and airflow without moving the setup.
Workflow diagram
Procedure
- Define the observation period, sampling and operating state.
- Record angle, temperature and time with the setup fixed.
- Mark power, operator and ventilation events.
- Inspect trends, short-term noise and correlations without assuming causation.
- Repeat observations to establish usable stability conditions and document any compensation or waiting policy.
Calculation
Relative drift d(t)=θ(t)−θ(t₀). A fitted time slope may be reported in arcsec/hour together with duration and operating conditions.
Illustrative example
Illustration: a change from 0.1″ to 0.4″ in one hour gives a 0.3″ endpoint change; two points do not prove linear drift.
Limitations
- Do not prescribe one universal warm-up time without evidence.
- Short-term averaging does not automatically remove slow drift.
Reporting
Retain raw data, datum definition, temperature, repeatability, algorithm and uncertainty. Illustrations are not test results.