Home / Printing press roller parallelism: the geometry behind register error
CASE 25 · TEACHING DATA

Printing press roller parallelism: the geometry behind register error

Register errors are often blamed on tension control, yet non-parallel roller axes shift the image linearly across the web. Measure parallelism first.

ObjectImpression cylinder, face length 1400 mm
InstrumentAutocollimator with bases on both sides
Readings4.4″ across the face
CriterionParallelism ≤ 0.03 mm over the web width
h_i = sum L tan(theta)

Why register shifts linearly

Non-parallel axes change contact pressure and surface speed across the face, so register drifts from one side to the other.

Tension adjustment changes overall stretch, not the cross-web gradient.

Conversion

Δh = 0.030 mm, slightly above the 0.03 mm criterion.

Over a 1 020 mm web the same angle gives 0.022 mm, still inside the 0.05 mm register allowance — hence "fails parallelism, passes print".

How to state the conclusion

Report both the full-face and the web-width figure, and cite which document defines the criterion.

Case workflow, animated

Image point (x, y)αX for pitch, Y for yawθx = x/2f, θy = y/2fSchematic · not to scale · not measured data
1 / 4
  1. Confirm the task and datum
  2. Lay out stations by span
  3. Acquire readings and convert
  4. Evaluate and verify

The animation follows the station layout and calculation order of this case. Values are teaching data.

ConclusionFull-face 0.030 mm slightly exceeds 0.03 mm, while the 1 020 mm web sees 0.022 mm, inside registration allowance.

Teaching data. Use the press maker specification; the example shows why the reference width matters.