Wafer stage guideway and flatness
Straightness, flatness and thermal stability of wafer, exposure and precision XY stages.
Instruments
- High-resolution autocollimator or interferometer
- Granite datum and vibration isolation
- Temperature and airflow monitoring
Procedure
- Confirm cleanliness, isolation and vibration limit.
- Stabilise thermally and establish the datum, avoiding hand heat.
- Measure straightness in both planes and table flatness.
- Run multi-point repeatability and reversal, logging thermal drift.
- Report per-axis error, flatness and drift curves.
Calculation
Straightness by angle integration or direct displacement; flatness on one datum; drift as reference-point motion over time.
Reporting
- Isolation and airflow stable
- Time and temperature logged
- Two planes separate
- Drift reported separately
Limitations
Sub-micron results depend on support, airflow and operator position; measure in the service state.
Understand the measurement step by step
- 01
Confirm cleanliness, isolation and vibration limit.
- 02
Stabilise thermally and establish the datum, avoiding hand heat.
- 03
Measure straightness in both planes and table flatness.
- 04
Run multi-point repeatability and reversal, logging thermal drift.
- 05
Report per-axis error, flatness and drift curves.
- Step 1 · Confirm cleanliness, isolation and vibration limit.
- Step 2 · Stabilise thermally and establish the datum, avoiding hand heat.
- Step 3 · Measure straightness in both planes and table flatness.
- Step 4 · Run multi-point repeatability and reversal, logging thermal drift.
- Step 5 · Report per-axis error, flatness and drift curves.
Teaching animation: follow the stations and reference directions. Do not read dimensions or angles from the schematic.
Worked 300 mm stage straightness and thermal drift example
Teaching example with assumed data, not a customer project, instrument accuracy claim or acceptance result.
Example task
Teaching example: a 300 mm travel split into five bays, angles integrated into straightness, with a reference point tracking thermal drift. Synthetic data.
Setup and preparation
Confirm cleanliness, isolation and vibration limit.
Stabilise thermally and establish the datum, avoiding hand heat.
Measure straightness in both planes and table flatness.
Run multi-point repeatability and reversal, logging thermal drift.
Example data and records
Bay angles in arc-seconds: +0.3, +0.2, -0.1, -0.2, -0.2 over 60 mm bays.
Accumulated height: 0.087, 0.145, 0.116, 0.058, 0 µm.
Reference point drifted 0.05 µm over 2 h, used to separate thermal effects.
Calculation steps
Δh = 0.087 µm; the accumulated range 0.145 µm is the straightness over this travel. Drift is reported separately and never subtracted from the profile.
Interpretation
Straightness is about 0.145 µm concentrated in the first bays, suggesting mounting or preload at the entry; drift is far smaller and not the cause.
Verification and pitfalls
Airflow and acoustic vibration disturb sub-micrometre readings; switch off non-essential equipment.
Let the sensor reach stage temperature; measuring on a cold start adds thermal error.
Keep straightness, flatness and angular error separate.