Rotary table and indexing head calibration
Full-circle indexing error and repeatability of NC tables, indexing heads and multi-tooth tables for acceptance, compensation and repair verification.
Instruments
- Polygon mirror or high-grade circular encoder
- 0.1″ autocollimator
- Centring fixture and temperature log
Procedure
- Confirm polygon faces, eccentricity and start angle; define stations.
- Mount the polygon coaxially and align the autocollimator over the full circle.
- Index face by face and record table and autocollimator readings.
- Compute indexing error and closure; run both directions to separate backlash.
- Build a compensation table if required and re-measure.
Calculation
Error = table reading − nominal polygon angle − zero offset. Closure checks consistency.
Reporting
- Polygon certificate valid
- Eccentricity checked
- Both directions acquired
- Closure within limit
Limitations
Polygon grade sets the floor; radial and axial runout are separate.
Understand the measurement step by step
- 01
Confirm polygon faces, eccentricity and start angle; define stations.
- 02
Mount the polygon coaxially and align the autocollimator over the full circle.
- 03
Index face by face and record table and autocollimator readings.
- 04
Compute indexing error and closure; run both directions to separate backlash.
- 05
Build a compensation table if required and re-measure.
- Step 1 · Confirm polygon faces, eccentricity and start angle; define stations.
- Step 2 · Mount the polygon coaxially and align the autocollimator over the full circle.
- Step 3 · Index face by face and record table and autocollimator readings.
- Step 4 · Compute indexing error and closure; run both directions to separate backlash.
- Step 5 · Build a compensation table if required and re-measure.
Teaching animation: follow the stations and reference directions. Do not read dimensions or angles from the schematic.
Worked 12-face polygon index error with closure check
Teaching example with assumed data, not a customer project, instrument accuracy claim or acceptance result.
Example task
Teaching example: the table is compared with a 12-face polygon every 30°; the sum of all deviations should close to zero and is used as a consistency check.
Setup and preparation
Confirm polygon faces, eccentricity and start angle; define stations.
Mount the polygon coaxially and align the autocollimator over the full circle.
Index face by face and record table and autocollimator readings.
Compute indexing error and closure; run both directions to separate backlash.
Example data and records
Index deviations in arc-seconds: +0.0, +0.4, -0.3, +0.2, -0.1, +0.5, -0.4, +0.1, -0.2, +0.3, -0.5, +0.0.
Closure 0″, peak-to-peak 1″.
Polygon grade and autocollimator resolution are logged separately; together they set the floor.
Calculation steps
Index error = table reading − nominal polygon angle − start zero; the twelve terms close to 0″. A non-zero closure means a zero drift or a reading error.
Interpretation
Peak-to-peak index error is 1″ with a near-sinusoidal pattern, typical of eccentric mounting.
Verification and pitfalls
Polygon and table must be coaxial; eccentricity appears as the first harmonic.
Read each position in both directions to separate backlash from true index error.
Temperature changes expand polygon and table differently; log it during long runs.