Home / Volumetric diagonal test
MEASUREMENT METHOD

Volumetric diagonal test

Screen volumetric accuracy after installation, relocation or repair and decide whether per-axis tests are needed.

Principle

A body diagonal combines positioning, straightness and squareness of three axes; comparing four diagonals separates directional terms.

Preparation

Confirm travel, diagonal directions and optical access; clean the table and floor and warm up the machine.

Workflow diagram

01Select diagonals02Set optical path03Acquire each04Compute metrics05Per-axis follow-up

Procedure

  1. Select four body diagonals and station distribution per ISO 230-6.
  2. Set up the interferometer or step gauge with a clear path.
  3. Acquire each diagonal and log environmental data.
  4. Compute diagonal deviation and volumetric metrics.
  5. Investigate out-of-tolerance directions with per-axis straightness and squareness.

Calculation

Compute diagonal deviation per ISO 230-6; the volumetric metric is usually the envelope of the four maxima.

Illustrative example

Schematic · not measured data
Schematic: four body diagonals and directional spread.

Example: maxima of 18 and 6 μm across diagonals indicate a directional term to investigate.

Limitations

  • A diagonal error does not identify a single axis.
  • Long paths need refractive-index compensation.
  • One diagonal misses directional effects.

Reporting

Retain raw data, datum definition, temperature, repeatability, algorithm and uncertainty. Illustrations are not test results.

STEP BY STEP

Understand the measurement step by step

Principle schematic · not to scale
1 / 5
  1. Step 1 · Select four body diagonals and station distribution per ISO 230-6.
  2. Step 2 · Set up the interferometer or step gauge with a clear path.
  3. Step 3 · Acquire each diagonal and log environmental data.
  4. Step 4 · Compute diagonal deviation and volumetric metrics.
  5. Step 5 · Investigate out-of-tolerance directions with per-axis straightness and squareness.

Teaching animation: follow the stations and reference directions. Do not read dimensions or angles from the schematic.

WORKED EXAMPLE

Worked body-diagonal example in a 800×600×500 mm volume

Teaching example with assumed data, not a customer project, instrument accuracy claim or acceptance result.

Example task

Teaching example: four body diagonals inside a 800×600×500 mm volume measured by length, following ISO 230-6, to locate the weak direction. Synthetic data.

Setup and preparation

  • Select four body diagonals and station distribution per ISO 230-6.

  • Set up the interferometer or step gauge with a clear path.

  • Acquire each diagonal and log environmental data.

  • Compute diagonal deviation and volumetric metrics.

Example data and records

  • Nominal diagonal length 1118.034 mm.

  • Deviation of the four diagonals in µm: +2.0, -1.0, +1.5, -0.5.

  • Ambient 20.4 °C, material temperature compensation enabled and logged.

Calculation steps

L = √(a²+b²+c²) = 1118.034 mm; the volumetric index is the range of the four deviations: 3 µm. It does not replace the individual ISO 230-1 geometric tests.

Interpretation

A large range points to one diagonal; split it into the single-axis straightness and squareness tests instead of adjusting compensation parameters.

Verification and pitfalls

  • Measure all four diagonals at the same feed, direction and warm-up state.

  • Check beam obstruction; long diagonals are the first to be blocked by fixtures.

  • Log the compensation state; never mix compensated and raw data.