MICROCRE / OPTICAL METROLOGY400-812-7833  ·  SHANGHAI
研润 WHW
MICROCREPRECISION IN EVERY ANGLE
Home / Rotary Positioning & Reversal
MEASUREMENT METHOD

Rotary Positioning & Reversal

Evaluate several rotary positions in both approach directions, separating position error, repeatability and reversal.

Principle

The reference supplies the large angle; the instrument reads a small residual within range.

Preparation

Select a calibrated polygon or angular generator; define positions, cycles, settling and direction.

Workflow diagram

01Verify stable mounting of the rotary stage and referen02Approach each position forward and record residual sma03Repeat in reverse04Apply reference corrections and obtain position deviat05Evaluate directional means

Procedure

  1. Verify stable mounting of the rotary stage and reference.
  2. Approach each position forward and record residual small angles.
  3. Repeat in reverse, preserving the approach-direction record.
  4. Apply reference corrections and obtain position deviations.
  5. Evaluate directional means, dispersion and reversal at each position.

Calculation

At each position, Bᵢ=mean(eᵢ,forward)−mean(eᵢ,reverse). Retain raw values and dispersion separately; use the specified standard for formal metrics.

Illustrative example

Schematic · not measured data
Principle and workflow schematic. Numbers are worked examples, not customer or instrument test results.

Illustration: mean deviations +0.6″ forward and +0.2″ reverse give +0.4″ directional difference.

Limitations

  • One cycle or one direction cannot separate all repeatability and reversal effects.
  • An uncalibrated polygon is not an error-free reference.

Reporting

Retain raw data, datum definition, temperature, repeatability, algorithm and uncertainty. Illustrations are not test results.

Method reference ↗ Method reference ↗ Method reference ↗ 
STEP BY STEP

Understand the measurement step by step

Principle schematic · not to scale
1 / 4
  1. Set angular targets · Define targets at 0°, 90°, 180° and 270°, with reference-polygon or reference-table corrections already included in the inputs.
  2. Record both directions · Approach each position in one direction, settle and record; then repeat from the opposite direction using consistent approaches.
  3. Subtract matched positions · Pair forward and reverse deviations at the same target and calculate signed differences, retaining them separately from averages.
  4. Report separately · Show the bidirectional mean and difference at each position; an actual test must use the cycles, start positions and indicators required by its chosen standard.

Teaching animation: follow the stations and reference directions. Do not read dimensions or angles from the schematic.

WORKED EXAMPLE

Bidirectional comparison at four rotary positions

Teaching example with assumed data, not a customer project, instrument accuracy claim or acceptance result.

Example task

Teaching demonstration, not a real customer project: corrected deviations at four positions illustrate direction-dependent differences. They are arithmetic inputs, not a calibration certificate or accuracy claim.

Setup and preparation

  • Define targets at 0°, 90°, 180° and 270°, with reference-polygon or reference-table corrections already included in the inputs.

  • Approach each position in one direction, settle and record; then repeat from the opposite direction using consistent approaches.

  • Pair forward and reverse deviations at the same target and calculate signed differences, retaining them separately from averages.

  • Show the bidirectional mean and difference at each position; an actual test must use the cycles, start positions and indicators required by its chosen standard.

Example data and records

  • Target positions: [0,90,180,270]°.

  • Corrected forward deviations: [+0.2,+0.8,−0.3,+0.1] arcsec; reverse: [+0.1,+0.4,−0.2,0] arcsec.

  • Direction differences: [+0.1,+0.4,−0.1,+0.1] arcsec.

Calculation steps

Bi=ei_forward−ei_reverse; max|Bi|=0.4 arcsec at 90°. The bidirectional mean there is (0.8+0.4)/2=0.6 arcsec.

Interpretation

The largest directional difference in this dataset is 0.4 arcsec at 90°. This arithmetic indicator does not replace a complete standard-defined angular positioning or repeatability evaluation.

Verification and pitfalls

  • Pair identical target positions rather than blindly subtracting row numbers.

  • Treat reference-face corrections and mounting eccentricity according to the actual model.

  • One four-position forward/reverse run neither maps all errors nor establishes repeatability.