Principle
The normals of a regular polygon's faces are nominally spaced at 360°/n. The polygon is mounted coaxially on the table under test while the autocollimator stays fixed on one face. After each nominal rotation of 360°/n the next face arrives and the autocollimator reads its angular change from the optical axis, the table's index deviation at that position. The polygon's own face errors superpose on the table error; the closure condition (deviations summing to zero) separates them, and rigorous procedures use cross-calibration permutations or corrections from the polygon's certificate.
Preparation
Confirm the polygon face count (commonly 12 or 24) matches the table's nominal indexing and check the polygon certificate. Mount the polygon coaxially within the specified tolerance and clean the faces. Set the autocollimator at the face centre height and check return light over the full circle. Ensure stable temperature and no draughts.
Workflow diagram
Procedure
- Mount and align the polygon, then rotate the table to check return light and reading range on every face.
- Zero on face 1 and record the initial autocollimator reading.
- Step the table by the nominal index angle, recording the table indication and autocollimator reading at each face.
- Complete the circle and return to face 1 to check the closure error.
- Separate table errors from polygon face errors using the closure condition, or apply certificate corrections.
- Report index errors, corrections and uncertainty with the calibration conditions.
Calculation
Per-position deviation e_i=A_i−A_1 (autocollimator reading difference with sign). Closure requires Σe_i=0; a residual indicates a faulty reading. Ignoring polygon errors, the table index error is e_i; rigorously e_i=t_i+p_i with table error t_i and polygon adjacent-face error p_i, separated by permutation comparison or certificate data. Sign conventions must be unified before reporting.
Illustrative example
Teaching data, not measured: calibrating a table with a 12-face polygon, the first three reading differences are +0.8″, −1.2″ and +0.4″. Neglecting polygon errors and assuming closure, these are the index errors at those positions, and the full-circle deviations must sum to zero. The numbers only illustrate the arithmetic and closure check.
Limitations
- Polygon eccentricity drifts the return image across faces and mixes mounting error into readings.
- Ignoring the polygon's own face errors can be commensurate with the error under test.
- A sign inversion applies corrections the wrong way and voids the whole calibration.
Reporting
Retain raw data, datum definition, temperature, repeatability, algorithm and uncertainty. Illustrations are not test results.