Principle
The reflector follows platform orientation; orthogonal channels resolve rotations in defined planes.
Preparation
Define machine coordinates, instrument-axis mapping, mirror orientation, stations and operating conditions.
Workflow diagram
Procedure
- Define the machine rotations represented by pitch and yaw.
- Mount the reflector rigidly and verify stability.
- Record both angular channels at stable travel stations.
- Subtract the agreed reference angles and plot angle versus position.
- Repeat forward and reverse travel and inspect coupling and repeatability.
Calculation
Δθx(x)=θx(x)−θx(x₀), Δθy(x)=θy(x)−θy(x₀). Report the reference, ranges and reversal differences.
Illustrative example
Illustration: 0.8″ forward and 0.5″ reverse at one station differ by 0.3″; this is not a complete positioning-accuracy result.
Limitations
- Two angular channels do not automatically measure roll.
- Quasi-static samples do not establish high-speed dynamic performance.
Reporting
Retain raw data, datum definition, temperature, repeatability, algorithm and uncertainty. Illustrations are not test results.