Principle
The evaluation reference is an ideal geometric element constructed from the data, not the physical measurement datum. The minimum-zone method searches two parallel lines or planes that contain all points with minimum separation. For a straightness profile the criterion is alternating contacts of high and low points on the reference; for a plane it is a crossing pattern of three high (low) and one low (high) points. Numerically it is a linear-programming or rotation-search problem: rotate and translate the reference to minimise the peak-to-valley distance.
Preparation
Prepare the integrated, unified point data and check span, numbering and sign conventions. Select a constrained-optimisation tool or verified evaluation program and validate it on simple cases first. Keep a copy of the raw data so the evaluation is reproducible.
Workflow diagram
Procedure
- Import the unified coordinates and heights; verify completeness and consistent units.
- Start from a least-squares line or plane and compute residuals and the peak-to-valley distance.
- Adjust rotation and translation parameters iteratively to narrow the containment band.
- At the minimum band, verify that the contact points satisfy the alternating (crossing) criterion.
- Record the optimum reference parameters, contact points and the minimum-zone value.
- Compare with endpoint and least-squares results and report the conclusion with reproducibility information.
Calculation
With reference z=a·x+b (plane: z=a·x+b·y+c), residuals are ri=zi−(a·xi+b). The minimum-zone problem is min over a,b of [max(ri)−min(ri)], solvable as linear programming. In manual rotation, Δa is chosen so that the current highest point and the runner-up residual become equal. The minimum-zone value never exceeds the peak-to-valley value from any other reference and is the smallest admissible form-error value.
Illustrative example
Teaching data, not measured: five equally spaced heights [0, −0.5, +1.0, 0, +0.3] μm. The endpoint reference gives 1.5 μm; rotating the reference to a slope of 0.25 μm per station narrows the band to 1.25 μm with low-high-low alternating contacts. The example shows that the minimum-zone value is the smallest among references.
Limitations
- The minimum-zone value holds only for the given point set; sparse sampling underestimates the error.
- Without convergence and the alternating-contact criterion the band is not a true minimum zone.
- Passing off a least-squares result scaled by an empirical factor as a minimum-zone value.
Reporting
Retain raw data, datum definition, temperature, repeatability, algorithm and uncertainty. Illustrations are not test results.