Home / Flatness by the Diagonal Method
MEASUREMENT METHOD

Flatness by the Diagonal Method

Evaluate the flatness of rectangular surfaces such as surface plates and machine tables. The diagonal method ties opposite corners into direct measurement chains, avoiding error accumulation around the perimeter. It suits plates on which a reflector base can traverse the whole area.

Principle

Along each measurement line the autocollimator records small tilts of the reflector base, and the angles are integrated into heights with Δh=L·tanθ. The two diagonals are measured first; equating their endpoints defines an interim reference plane. Perimeter and intermediate parallel sections are then measured and transformed onto that plane. The resulting point cloud supports flatness evaluation against the agreed reference.

Preparation

Define the surface size, tolerance, and a layout of diagonals, perimeter and intermediate sections with section spacing equal to an integer multiple of the base span. Clean the surface and record the measured support span of the reflector base. Verify autocollimator calibration, return-beam brightness and thermal stability; avoid one-sided sunlight or draughts.

Workflow diagram

01Plan the grid02Measure diagonals03Measure sections04Unify to datum05Evaluate flatness06Report with uncertainty

Procedure

  1. Draw the layout so that section spacing equals the base span; number every station on both diagonals, the perimeter and the intermediate sections.
  2. Measure the diagonals first: step the reflector base along each diagonal, record stable readings and integrate to the end-point height differences.
  3. Define the reference plane by equating the diagonal endpoints, then measure the perimeter and intermediate sections.
  4. Unify each section to the datum: align it at the intersections with the diagonals and shift the whole section accordingly.
  5. Check the closure error of points shared by crossing sections; re-measure the offending segments if it is excessive.
  6. Assemble all point heights, evaluate flatness against the agreed reference and report the result with uncertainty.

Calculation

Per station: θrad=θ″·π/648000 and Δh=L·tan(θrad), with L the measured span in mm (×1000 for μm). Equating diagonal endpoints fixes the datum; each section is shifted by its intersection offset: z′i=zi+δ. Flatness is then evaluated against the agreed reference such as the minimum zone.

Illustrative example

Schematic · not measured data
Schematic: two diagonals and parallel sections; arrows show sequence, data are illustrative.

Teaching data, not measured: on an 800 mm × 500 mm plate with L=100 mm a diagonal has about 10 stations. If the integrated end difference is +2.4 μm, the datum correction is −1.2 μm at each end. A section whose intersection mismatch is +0.3 μm is shifted by −0.3 μm. These numbers only illustrate the workflow.

Limitations

  • If the measured diagonals deviate from the geometric ones, the closure relations break down.
  • A base span that is not an integer fraction of the section spacing misaligns the integrated nodes.
  • Treating the interim-datum result as the minimum-zone flatness value.

Reporting

Retain raw data, datum definition, temperature, repeatability, algorithm and uncertainty. Illustrations are not test results.

Method reference ↗ Method reference ↗ 
STEP BY STEP

Understand the measurement step by step

Principle schematic · not to scale
1 / 6
  1. Top view of the plate: the nine stations light up in a 3 × 3 grid and both diagonals are drawn in orange.
  2. The reflector base steps heel-to-toe along diagonal AC while the cross-hair swings with each tilt.
  3. Below the view the two angular readings accumulate into a height polyline that returns to zero.
  4. The four corner points are pulled onto one plane and the datum plane appears above the plate.
  5. The cross sections unfold from the centre point and four height values drop into place.
  6. All nine stations merge into a point cloud with a 3.3 μm band between the highest point (0, 200) and the lowest.

Teaching animation: follow the stations and reference directions. Do not read dimensions or angles from the schematic.

WORKED EXAMPLE

Flatness of a 400 mm × 400 mm granite surface plate by the diagonal method

Teaching example with assumed data, not a customer project, instrument accuracy claim or acceptance result.

Example task

A 400 mm × 400 mm granite plate is checked in a metrology room held at 20.3 °C. The plate rests on three adjustable supports and was cleaned with alcohol 30 min earlier. An autocollimator with 0.01″ resolution sits on a rigid stand beside the plate, and the reflector base span equals half the diagonal, 282.8 mm.

Setup and preparation

  • Check the surface with a grade 00 straight edge and feeler gauge, then mark a 3 × 3 grid at 200 mm pitch and number the nine stations.

  • Mark both diagonals AC and BD plus the cross sections through the centre; the diagonal spans are 282.8 mm and the cross-section spans 200 mm.

  • Clamp the autocollimator on the external stand with the axis nearly parallel to the surface, incidence below 3′, and do not touch the stand during the run.

  • The measured support span of the reflector base is 282.8 mm; place it heel-to-toe along each diagonal with no gap under the feet.

  • Warm up for 30 min and zero the instrument; record 20.3 °C and 46 % RH.

Example data and records

  • Diagonal AC, segment 1 (0.0→282.8 mm): +1.80″

  • Diagonal AC, segment 2 (282.8→565.6 mm): -1.80″

  • Diagonal BD, segment 1 (0.0→282.8 mm): +1.72″

  • Diagonal BD, segment 2 (282.8→565.6 mm): -1.72″

  • Mid section y=200 mm (x=200→0 mm): +0.90″

  • Mid section y=200 mm (x=200→400 mm): -0.80″

  • Mid section x=200 mm (y=200→0 mm): -1.30″

  • Mid section x=200 mm (y=200→400 mm): +0.60″

Calculation steps

Scale factors: kd=282.8×π/648000×1000=1.371 μm per arcsec and km=200×π/648000×1000=0.970 μm per arcsec. Diagonal AC height steps Δh=+2.47 → -2.47 μm, cumulative +0.00 → +2.47 → +0.00 μm, closing at 0.00 μm. Diagonal BD height steps Δh=+2.36 → -2.36 μm, cumulative +0.00 → +2.36 → +0.00 μm, end closure +0.00 μm. Equating the four corner points of both diagonals defines the datum plane, so the cumulative heights are already heights above that plane and no linear correction is needed. The centre is measured twice: AC gives +2.47 μm and BD gives +2.36 μm, a closure of 0.11 μm; the mean +2.41 μm is adopted. The cross sections are integrated outward from the centre: +3.29 μm at x=0, +1.64 μm at x=400, +1.15 μm at y=0 and +2.99 μm at y=400. Of the nine stations the highest is +3.29 μm at (0, 200) and the lowest +0.00 μm at (0, 0), so the flatness is f=3.29−(+0.00)=3.29 μm.

Interpretation

Flatness by the diagonal method: 3.3 μm. This passes the 8 μm grade 1 tolerance for a 400 mm × 400 mm plate with 4.7 μm margin and also stays inside the 4 μm grade 0 tolerance with 0.7 μm to spare.

Verification and pitfalls

  • Both diagonals must close on their end points: AC closes at 0.00 μm and BD at 0.00 μm here, confirming the sign convention and integration direction.

  • The centre is measured on both diagonals with a closure of 0.11 μm; investigate the central stations and the base seating above 0.5 μm.

  • The cross-section ends must land on already measured corner points; a mismatch above 0.3 μm after conversion means a wrong span or a misplaced station.

  • Repeat each station twice and require a difference of no more than 0.05″; re-clean and re-seat the base if it is larger.

  • Pause when the temperature moves by more than 0.5 °C or when air movement makes the cross-hair flicker, then resume from the last segment.