Granite & Precision Bases
Granite plates, straightedges, equipment bases and support surfaces
Methods
- Grid flatness
- Line straightness
- Support-state reproduction
Procedure
- Define supports
- Stabilize temperature
- Measure the grid
- Check node closure
- Evaluate and verify
Support changes can alter form; report support conditions and sampling coverage.
Typical measurement scenarios in this industry
Each scenario gives the object, tolerance range and method, ready to take to the selector.
Grade 000 granite plate inspection
A 1000×1500 mm plate on a 9×7 grid after 24 h soak: rough scan for high points, then fine measurement with a minimum-zone plane.
Metrology autocollimator series↗Lapping correction closed loop
Measure, map high points, plan the lapping path, measure again. Remove 1–2 μm per pass to avoid destroying the datum.
Metrology autocollimator series↗Granite straight edge and square
Measure the working face by the pitch method, supporting at the two Bessel points to minimise self-weight sag.
Metrology autocollimator series↗Granite machine bed levelling
Granite has low expansion but is brittle: support evenly at many points, then re-check after 12 h for creep.
Metrology autocollimator series↗Plate certificate and uncertainty
A certificate must state grid, support, temperature, evaluation method and U. A bare 'pass' cannot be reproduced or defended.
Metrology autocollimator series↗Verification of granite V-blocks and straightedges
Periodic verification of granite V-blocks and straightedges for parallelism and straightness; autocollimator and optical-flat methods cross-check each other.
Metrology autocollimator series↗Granite square perpendicularity
300–1000 mm, typically 3–5 μm; reversal separates instrument and part error.
General-purpose electronic autocollimators↗Straightedge straightness
1–3 m straightedges; evaluate against end points and correct for sag.
General-purpose electronic autocollimators↗Understand the measurement step by step
- 01
Define supports
- 02
Stabilize temperature
- 03
Measure the grid
- 04
Check node closure
- 05
Evaluate and verify
- Mark supports · Mark support positions, pads and base orientation; record temperature and settling state and establish a reference that does not move with the base.
- Record state A · Measure four fixed marks in support state A and retain their coordinates and zero-transfer records.
- Reposition to B · Lift and reposition under the approved handling procedure, restore support state B, allow settling and remeasure against the same external reference.
- Differences and grid · Compute B−A at each point, inspect support contact near changed locations, and use a complete grid for final surface evaluation.
Teaching animation: follow the stations and reference directions. Do not read dimensions or angles from the schematic.
Reproducing granite-base support after repositioning
Teaching example with assumed data, not a customer project, instrument accuracy claim or acceptance result.
Example task
Teaching demonstration, not a real customer project. All numbers are constructed examples. Check whether marked points change relative to the same external height reference after the base is relocated. The aim is support reproducibility, not a four-point claim about whole-surface flatness.
Setup and preparation
Mark support positions, pads and base orientation; record temperature and settling state and establish a reference that does not move with the base.
Measure four fixed marks in support state A and retain their coordinates and zero-transfer records.
Lift and reposition under the approved handling procedure, restore support state B, allow settling and remeasure against the same external reference.
Compute B−A at each point, inspect support contact near changed locations, and use a complete grid for final surface evaluation.
Example data and records
State A heights in μm: [0,2,−1,1].
State B heights in μm: [0,1,−2,1].
Point differences B−A in μm: [0,−1,−1,0].
Calculation steps
Δzᵢ=zBᵢ−zAᵢ; max|Δzᵢ|=1 μm and the difference range is 0−(−1)=1 μm.
Interpretation
The maximum absolute change at the four sampled positions is 1 μm. Its significance depends on reference stability and uncertainty; it does not establish 1 μm flatness of the entire base.
Verification and pitfalls
Independently verify the external reference: its motion could mimic base motion.
Keep support points, pads and fastening consistent, recording unavoidable changes.
Complete thermal settling and full-grid verification before judging the final installation state.