Display & Electronics Manufacturing
Glass-substrate stages, inspection guides and precision placement fixtures
Methods
- Table datum relationships
- Stage pitch and yaw
- Fixture parallelism
Procedure
- Define support and vacuum state
- Establish references
- Record stage orientation
- Adjust and lock
- Recheck repeatability
Loaded deformation of thin or flexible glass needs separate measurement; base geometry is not the complete substrate form.
Typical measurement scenarios in this industry
Each scenario gives the object, tolerance range and method, ready to take to the selector.
Large panel and stage flatness
Substrates of 1.5–3 m need 10–50 μm over the whole area. Measure in tiles, then stitch with a common datum and overlap.
Metrology autocollimator series↗Laminator upper and lower plate parallelism
OCA lamination needs 5–20 μm between plates. Check four corners plus centre, and again under air pressure.
Wide-field electronic autocollimators↗Laser lift-off and cutting stage straightness
Stages of 1–2 m turn straightness error into kerf offset. Measure at stable temperature with air bearings on.
Laser autocollimators↗Mask to substrate angle
In proximity exposure the mask-to-substrate angle shifts line width. Two points give the angle, converted through the exposure gap.
Metrology autocollimator series↗AOI station repeatability validation
Thirty repeats on one reference panel reveal false-call rate and angle drift; insufficient warm-up is the usual early culprit.
Zhongji high-speed dynamic autocollimators↗Parallelism of coating-line conveyor rolls
Roll parallelism on display coating lines drives substrate tracking; compare roll by roll with autocollimator and penta prism at 0.02–0.05 mm/m.
Zhongji automatic measurement series↗Recommended instruments


Zhongji high-speed electronic autocollimator accuracy 1″

Zhongji high-speed electronic autocollimator accuracy 2″
Understand the measurement step by step
- 01
Define support and vacuum state
- 02
Establish references
- 03
Record stage orientation
- 04
Adjust and lock
- 05
Recheck repeatability
- Fix loading state · Fix support conditions and measurement direction; record whether glass is loaded, the vacuum setting and temperature.
- Record vacuum off · Mount the reflector on a rigid stage datum and record a stable angle and reference zero with vacuum off.
- Apply vacuum and reread · Apply the specified vacuum under the equipment procedure, wait for pressure and mechanical settling, read the same component and switch vacuum off to check return.
- Separate direction and form · Convert the state difference to equivalent directional displacement at the stated lever arm while checking local substrate form independently.
Teaching animation: follow the stations and reference directions. Do not read dimensions or angles from the schematic.
Display-substrate stage direction before and after vacuum loading
Teaching example with assumed data, not a customer project, instrument accuracy claim or acceptance result.
Example task
Teaching demonstration, not a real customer project. All numbers are constructed examples. Compare one angular component of a rigid display-panel fixture with vacuum off and on. Thin-glass bending requires independent surface measurement and cannot be inferred directly from stage angle.
Setup and preparation
Fix support conditions and measurement direction; record whether glass is loaded, the vacuum setting and temperature.
Mount the reflector on a rigid stage datum and record a stable angle and reference zero with vacuum off.
Apply the specified vacuum under the equipment procedure, wait for pressure and mechanical settling, read the same component and switch vacuum off to check return.
Convert the state difference to equivalent directional displacement at the stated lever arm while checking local substrate form independently.
Example data and records
Vacuum-off angle 0.6″; vacuum-on angle 1.1″; difference 0.5″.
Lever arm for directional conversion: 150 mm.
No glass-form data are supplied in this example; glass deformation is not inferred.
Calculation steps
Δθ=1.1−0.6=0.5″; e=150×tan(0.5π/648000)×1000≈0.3636 μm.
Interpretation
The difference corresponds to about 0.364 μm equivalent directional displacement at 150 mm. It excludes stage translation, local suction deformation and glass warpage.
Verification and pitfalls
Record pressure settling time and loading conditions; do not combine different states as one test.
Verify attachment to a rigid datum rather than a flexible cover.
Acquire vacuum-release return and repeated loading cycles separately.